• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD INC. (COVINGTON) -1018233 BARD® SILASTIC® FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD INC. (COVINGTON) -1018233 BARD® SILASTIC® FOLEY CATHETER Back to Search Results
Catalog Number 33422
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 04/17/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that foley catheter been reported as defective, according to the patient the balloon burst inside and has to be hospitalized for 5 months, another instance was the balloon was left inside.They seek medical attention.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that foley catheter been reported as defective, according to the patient the balloon burst inside and has to be hospitalized for 5 months, another instance was the balloon was left inside.They seek medical attention.
 
Additional Manufacturer Narrative
The reported event was inconclusive because this investigation did not result in any additional findings and no sample was available for evaluation the reported event is addressed within the labeling as it states: visually inspect the product for any imperfections or surface deterioration prior to use.To deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If necessary, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.No additional actions can be taken at this time.Corrections: d, e.Udi format updated with available product information per gudid.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® SILASTIC® FOLEY CATHETER
Common Device Name
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key24628320
Report Number1018233-2026-01841
Device Sequence Number18441918
Product Code EZL
UDI-Device Identifier00801741020018
UDI-Public(01)00801741020018(11)241114(17)270731(10)NGJX4255
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K951103
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/05/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number33422
Device Lot NumberNGJX4255
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/12/2026
Supplement Date Received by Manufacturer05/05/2026
Initial Report FDA Received Date03/18/2026
Supplement Report FDA Received Date05/11/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/14/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other; Hospitalization;
Patient SexUnknown
-
-