| Model Number |
SIGTRSB60AXT |
| Medical Device Problem Code |
Misfire (2532)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
02/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D10 concomitant product: egiaushort, egiaushort endogia ultra univ sht stap, (lot # p5g1131).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a laparoscopic right upper lobe resection, the firing stopped midway.To resolve the issue, the handle was replaced, and then the firing was done using the same cartridge, but because it was not possible afterwards, the cartridge was also replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the reload was pre-fired and the interlock was engaged.Staples were protruding from the proximal end of the staple cartridge channel.It was reported that the device did not fire completely.The reported issue could not be confirmed.The most likely cause could not be established from the information available.This issue can occur if the firing button had been partially compressed and then the open button was pressed after pressing the green firing button.The evaluation detected an unreported condition: of device prefired.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: failure to completely fire the reload will result in an incomplete cut and/or incomplete staple formation, which may result in poor hemostasis and/or leakage.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|