• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LIGASURE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LIGASURE; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number LF1837
Medical Device Problem Codes Failure to Cut (2587); Difficult to Open or Close (2921); Intermittent Energy Output (4025)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/23/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported during the laparoscopic colectomy, the physician closed the device, sealed, then went to cut and release, but it would not cut and release from ileocolic vascular pedicle after 20 minutes of the procedure.The device's knife blade was extended.The device was released from the tissue by forcefully pulling the handles apart.There was no patient injury.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the device experienced a cutting issue, stopped working, and the jaws were difficult to open.The reported issues could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGASURE
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
COVIDIEN
6044 longbow court
boulder CO 80301 3299
Manufacturer (Section G)
COVIDIEN
6044 longbow court
boulder CO 80301 3299
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key24629100
Report Number1717344-2026-00400
Device Sequence Number13028865
Product Code GEI
UDI-Device Identifier10884521195219
UDI-Public(01)10884521195219(17)300619(10)51710088X(20)02
Combination Product (Y/N)N
Initial Reporter StateME
Initial Reporter CountryUS
PMA/510(K) Number
K162941
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/03/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberLF1837
Device Catalogue NumberLF1837
Device Lot Number51710088X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/17/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/24/2026
Supplement Date Received by Manufacturer05/20/2026
Initial Report FDA Received Date03/18/2026
Supplement Report FDA Received Date06/03/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/20/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-