| Catalog Number |
136532000 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Joint Dislocation (2374)
|
| Date of Event |
03/03/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that patient was revised from metal on metal pinnacle to constrained liner due to dislocations.Doi: (b)(6) 2007.Dor: (b)(6) 2026.Affected side: left knee.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional narrative: d6, h4: the information provided regarding the manufacturing date and implantation date are identified to be conflicting.Continued follow-up is being conducted and any additional information received will be reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The adverse symptoms alleged and product code reported is associated with the depuy metal on metal articulation.A complaint database search and/or device manufacturing (dhr) reviews will not be performed.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: a manufacturing record evaluation (mre) will not be performed since mom systems are obsolete.Added: d10 (concomitant).Corrected: d4 ( expiration date), (primary udi number), h4.
|
| |
|
Search Alerts/Recalls
|