Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id: 9735825, serial/lot #: (b)(6).H3, h6: the system was serviced in the field by a medtronic representative.The electromagnetic (em) interface was replaced to resolve the issue.Codes b01, c07, and d02 are applicable.The em interface was returned for evaluation.There was physical damage.The complaint was verified.The returned em interface's cable was torn in multiple spots with exposed wires.Despite the damage, the emitter did pass testing.Codes b01, c07, and d02 are applicable.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|