• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AVANOS MEDICAL INC. NEOMED NASOGASTRIC/OROGASTRIC FEEDING TUBES; DH NASOGASTRIC (NG) FEEDING TUBES (NEOMED AND NEOCONNECT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AVANOS MEDICAL INC. NEOMED NASOGASTRIC/OROGASTRIC FEEDING TUBES; DH NASOGASTRIC (NG) FEEDING TUBES (NEOMED AND NEOCONNECT) Back to Search Results
Model Number PFTS6.5P-NC
Medical Device Problem Codes Obstruction of Flow (2423); Pumping Problem (3016)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Malfunction
Event or Problem Description
Avanos medical inc.Received a single report that referenced different incidents, which were associated with separate units, involving different patients.This is the first of three reports.Refer to 3011270181-2026-00037 for the second report.Refer to 3011270181-2026-00038 for the third report.The customer reported that clinicians on the infant toddler floor experienced occlusion alarms when using kangaroo pumps with avanos 6.5 fr neomed feeding tubes (ref: pfts6.5p-nc) during infusion.The pump alarmed ¿occluded¿ and did not respond to troubleshooting.The issue occurred during use with pedialyte (thin liquid), and clinicians suspected a possible psi or pressure setting issue related to the feeding tube.Multiple events were reported with this tube size over the course of a week.The exact quantity is unknown.No patient injury was reported.
 
Additional Manufacturer Narrative
The actual complaint product was not returned for evaluation.A review of the device history record is not possible as no lot number was provided; therefore, the udi-pi is unavailable.A root cause could not be determined.All information reasonably known as of 17 mar 2026 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the reported event but is relaying the information provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEOMED NASOGASTRIC/OROGASTRIC FEEDING TUBES
Common Device Name
DH NASOGASTRIC (NG) FEEDING TUBES (NEOMED AND NEOCONNECT)
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
NINGBO TIANYI MEDICAL APPLIANCE CO., LTD
no.788,mozhi north rd
tourism resort, dongqian lake
ningbo zhejiang, cn 31512 1
CH   315121
Manufacturer Contact
tamara cardona
5405 windward parkway
alpharetta, GA 30004
4704485487
MDR Report Key24630567
Report Number3011270181-2026-00036
Device Sequence Number12933793
Product Code FPD
UDI-Device Identifier00350770001784
UDI-Public(01)00350770001784
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K082238
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 03/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPFTS6.5P-NC
Device Catalogue NumberPFTS6.5P-NC
Device Lot NumberUNKNOWN
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/18/2026
Initial Report FDA Received Date03/18/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
KANGAROO PUMPS
Patient SexUnknown
-
-