| Model Number |
P1190A |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/20/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The baxter technician found the siderail cable needed to be replaced.Unable to determine cause of impact that would tear off siderail from frame.Per the hillrom service manual the advanta 2 should be subject to an effective maintenance program.This will help make sure of a long, operative life for the advanta 2 bed.The preventative maintenance will help to reduce downtime due to excessive wear failures.An annual service of the bed is advised in order to maintain its characteristics and performance.Make sure the head and foot siderails are not bent or twisted.Make sure the latch mechanism operates correctly.An audible click must be heard.Remove the siderail cover, and make sure the mounting screws are tight.Inspect the cable routing for pinching, binding, and damage.Make sure all functions on the caregiver control operate correctly.Repair or replace the siderail as necessary.A search of the baxter maintenance records showed baxter performed preventative maintenance on this bed in aug 27, 2024.It is unknown if the facility performed any other preventative maintenance on this bed.The technician replaced the siderail cable assembly, control membrane and bracket to resolve the reported event.Based on this information, no further action is required.
|
| |
|
Event or Problem Description
|
|
On 20-feb-2026, a company representative - service personnel contacted technical service to report that advanta 2 frame (product code p1190a000091, serial number (b)(6)), had an issue, account stated siderail detached.This occurred after patient use.There was no patient/user injury, medical intervention, symptom, or adverse event reported.
|
| |
|
Search Alerts/Recalls
|
|
|