• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERPICC; PERIPHERALLY INSERTED CENTRAL CATHETER (PICC)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERPICC; PERIPHERALLY INSERTED CENTRAL CATHETER (PICC) Back to Search Results
Catalog Number 1173118
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/12/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.Each event reported to bd is evaluated and investigated in accordance with our complaint investigation procedures.The investigation process includes, but is not limited to, evaluation of the event details provided by the complaint facility, a review of complaint history, manufacturing records and risk document where applicable, and an evaluation of the subject device when available to identify potential contributing factors.The device has not been returned to the manufacturer for evaluation.
 
Event or Problem Description
It was reported that during an infusion procedure, a downstream occlusion alarm activated.When dressing was removed, blood was noted, and it was observed that the picc line had cracked at the bottom of the white bard component.The event occurred on (b)(6) 2026.It was reported as no apparent harm, though the patient experienced inconvenience, and the incident resulted in unexpected or prolonged care.No further information is provided.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The complaint of a split in catheter was confirmed.The product returned for evaluation was one 3 fr s/l powerpicc sv catheter.The investigation findings are consistent with damage accumulated through flexural fatigue.Flexural fatigue may occur due to cyclic kinking of the catheter tube in which physiological, placement, usage, and mechanical factors may gradually form a crack(s) in the catheter.The returned product sample was evaluated, and a split was observed just distal to the molded suture wings.The catheter split contained physical features associated with material fatigue, and the characteristics observed which supported this type of failure included: ¿ damage which was circumferentially aligned ¿ fracture edges which were rounded and polished due to repeated material wear ¿ 'c' shaped impressions leading into the fracture site which are consistent with material buckling due to movement which caused the fracture edges to be pressed together ¿ overall elliptical shape to the fracture cross-section (a result of repeated kinking of the tubing) an examination of the catheter structure revealed no potential damage/defect related to manufacture of the product.The damage location suggested that catheter securement, access and maintenance techniques may have contributed.This complaint will be recorded for future trending and monitoring purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPICC
Common Device Name
PERIPHERALLY INSERTED CENTRAL CATHETER (PICC)
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
veronica lee
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key24633976
Report Number3006260740-2026-01722
Device Sequence Number21123201
Product Code LJS
UDI-Device Identifier00801741086892
UDI-Public(01)00801741086892(11)250322(17)260731(10)REKN1517
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K102159
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/02/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2026
Device Catalogue Number1173118
Device Lot NumberREKN1517
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/02/2026
Supplement Date Received by Manufacturer05/01/2026
Initial Report FDA Received Date03/18/2026
Supplement Report FDA Received Date05/05/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/22/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-