U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G7 CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G7 CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 9500-161
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/09/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
Subsequent to the initial mdr, a correction is required.
 
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that an applicator deployment occurred.The sensor was inserted into the arm on (b)(6) 2025.Product was provided for investigation.An external visual inspection was performed and failed.An applicator deployment was not found within the investigation window.The allegation was not confirmed.However, a broken sensor wire was found in connection to the reported allegation.The probable cause of the broken sensor wire could not be determined.No injury or medical intervention was reported.
 
Additional Manufacturer Narrative
(b)(4).B5: describe event or problem ¿ correction.H2: type of follow up- correction.H6: component code- correction.
 
Event or Problem Description
Subsequent to the initial mdr, a correction is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G7 CONTINUOUS GLUCOSE MONITORING SYSTEM
Common Device Name
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence dr.
san diego CA 92121
Manufacturer Contact
jaime chavez
6340 sequence dr.
san diego, CA 92121
8582000200
MDR Report Key24634416
Report Number3004753838-2026-114785
Device Sequence Number8698217
Product Code QBJ
UDI-Device Identifier00386270003935
UDI-Public(01)00386270003935(11)250301(17)260831(10)1825085005
Combination Product (Y/N)N
PMA/510(K) Number
K213919
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/31/2026
Device Model Number9500-161
Device Catalogue NumberSTP-GT-002
Device Lot Number1825085005
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/22/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/18/2026
Supplement Date Received by Manufacturer03/23/2026
03/23/2026
Initial Report FDA Received Date03/18/2026
Supplement Report FDA Received Date03/23/2026
03/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/01/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
FIASP PEN.
Patient Age42 YR
Patient SexMale
-
-