• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 3015516266 OSTEOTOME HANDLE; ORTHOPEDIC HANDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 3015516266 OSTEOTOME HANDLE; ORTHOPEDIC HANDLE Back to Search Results
Catalog Number 17001A
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/03/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.
 
Event or Problem Description
It was reported that watson set 2 blade handle strike plate broke.There was no delay in the case.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees hat the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: d4 (lot), d9, h4.Corrected: d4 (primary udi).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: according to the information received, ¿it was reported that watson set 2 blade handle strike plate broke.There was no delay in the case.Product will be returned." the osteotome handle (lot.Pg352908) returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual inspection revealed that the strike plate was fractured and separated from the device; the fragment was returned for evaluation.Additionally, what appear to be impaction marks were observed throughout the proximal section of the shaft, below the strike plate, which is an area not intended for impactions.Fracture mechanism was due to the bending moment translated from impactions on the strike plate with those impactions occurring slightly off axis of the shaft of the instrument in essentially random directions of off axis impaction.The observed condition was identified as an end-of-life indicator, damage consistent with repeated use and servicing.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.A dimensional inspection and functional testing were not performed as it is not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the osteotome handle would contribute to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSTEOTOME HANDLE
Common Device Name
ORTHOPEDIC HANDLE
Manufacturer (Section D)
DEPUY IRELAND - 3015516266
loughbeg ringaskiddy co.
cork
EI 
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key24634662
Report Number1818910-2026-04951
Device Sequence Number8852457
Product Code LXH
UDI-Device Identifier00197169170016
UDI-Public(01)00197169170016(10)PG352908
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number17001A
Device Lot NumberPG352908
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/03/2026
Supplement Date Received by Manufacturer03/24/2026
04/09/2026
Initial Report FDA Received Date03/18/2026
Supplement Report FDA Received Date03/30/2026
04/10/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/05/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age80 YR
Patient SexMale
-
-