Updated fields: d8, d9, h2, h3, h6.This supplemental report is being submitted to provide the results of the final investigation.The device was returned to olympus for inspection and the reported failure was confirmed.The image flickers and has image blackout during angulation adjustment.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: based on the results of the investigation and historical data, reasonably known information suggests probable causes such as damage or deterioration of parts, environment problem like as dust/dirt/humidity, optical problem, overheating problem, and electrical problems like as signal loss or open circuit.The most probable cause of this complaint is expected or random component failure without any design or manufacturing issue.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|