• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. BRONCHOVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. BRONCHOVIDEOSCOPE Back to Search Results
Model Number BF-1TQ170
Medical Device Problem Codes Erratic or Intermittent Display (1182); No Display/Image (1183)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/27/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
 
Event or Problem Description
It was reported the bronchovideoscope exhibited a blackout during angulation adjustment.The issue occurred during a inspection for use.There were no reports of patient harm.
 
Additional Manufacturer Narrative
Updated fields: d8, d9, h2, h3, h6.This supplemental report is being submitted to provide the results of the final investigation.The device was returned to olympus for inspection and the reported failure was confirmed.The image flickers and has image blackout during angulation adjustment.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: based on the results of the investigation and historical data, reasonably known information suggests probable causes such as damage or deterioration of parts, environment problem like as dust/dirt/humidity, optical problem, overheating problem, and electrical problems like as signal loss or open circuit.The most probable cause of this complaint is expected or random component failure without any design or manufacturing issue.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
No additional information received from the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRONCHOVIDEOSCOPE
Common Device Name
BRONCHOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key24635525
Report Number9610595-2026-22439
Device Sequence Number18439759
Product Code EOQ
UDI-Device Identifier04953170342943
UDI-Public04953170342943
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K121959
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/08/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberBF-1TQ170
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/27/2026
Supplement Date Received by Manufacturer03/23/2026
Initial Report FDA Received Date03/18/2026
Supplement Report FDA Received Date04/08/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/18/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-