| Model Number |
CD3369-40C |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Unspecified Infection (1930); Pocket Erosion (2013)
|
| Date of Event |
03/10/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient presented to hospital with infection and skin erosion at device pocket.The cause of the infection was not identified.Echocardiography showed no evidence of vegetation noted on the leads.The implantable cardioverter defibrillator (icd), right atrial lead, right ventricular lead and left ventricular lead were explanted.The patient was in stable condition throughout the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
A device history record (dhr) review was performed and all required manufacturing processes and inspection steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.Review of the sterilization records confirmed normal sterilization cycles for the products.The product was returned and visual inspection was normal.Interrogation of the device revealed it was above elective replacement indicator (eri) when received.The cause of infection could not be traced to the device.
|
| |
|