| Model Number |
GL-885 |
| Medical Device Problem Codes |
Break (1069); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/23/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D10 concomitant product: gl-885, gl-885 polysorb* 5-0 und 75cm cv22 x36 (lot#d4a3062y); gl-885, gl-885 polysorb* 5-0 und 75cm cv22 x36 (lot#d4a3062y); medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, prior to use, upon inspection, two needle-suture pairs were found to be detached, and all three needles had absorbable suture that was brittle and broke with a slight pull.The product was not used on the patient; it was replaced to resolve the issue.
|
| |
|