(b)(4).Date sent: 3/19/2026.D4 batch #: unknown.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
(b)(4).Date sent: 4/16/2026.D4 batch #: a99433.Investigation summary : the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis determined that the returned device had the distal tip of the blade broken off and not returned, with the remaining blade portion scratched and showing evidence of contact with metal.During functional testing, an alert screen was displayed, which is likely due to blade damage.Disassembly verified that internal components showed no anomalies.Once minor blade damage has occurred, subsequent activations may increase the severity of the blade damage, resulting in activation issues such as failing the pre-run test and displaying alert screens like ¿blade error detected¿ or "relaxed pressure on blade," followed by a ¿replace instrument¿ screen.Continued usage of the damaged blade can result in a broken blade.Probable causes of blade damage include external contact during pre-op or general use, blade contact with other devices, staples, or clips during the procedure, or using any means other than the blade wrench to attach or detach the blade.As part of the quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch a99433, and no non-conformances were identified.
|