• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC ARROW PI PICC 1L 4FRX50CM W/80CM HYDRO NIT SWG; CATHETER, INTRAVASCULAR, THER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC ARROW PI PICC 1L 4FRX50CM W/80CM HYDRO NIT SWG; CATHETER, INTRAVASCULAR, THER Back to Search Results
Catalog Number PR-35041-HPHNM
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/26/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported "difficult picc insertion.Insertion ceased due to inability to advance catheteter.Attempted removal unsuccessful as catheter 'lodged' insitu.Xray ordered and picc visualised as being knotted.Patient referred for surgical review.Surgical team allegedly pulled on the catheter in attempt to remove it however were unsuccessful and in the process snapping the catheter leaving some catheter insitu.Patient will now be transported to tertiary hospital for vascular intervention.The patient's current condition is reported as "fine".Associate mdr numbers include: 9680794-2026-00204 and 9680794-2026-00221.
 
Additional Manufacturer Narrative
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW PI PICC 1L 4FRX50CM W/80CM HYDRO NIT SWG
Common Device Name
CATHETER, INTRAVASCULAR, THER
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL DE MEXICO S.A. DE C.V.
ave. washington 3701
colonia panamericana, chihuahua
chihuahua 31200
MX   31200
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key24639263
Report Number9680794-2026-00221
Device Sequence Number9895027
Product Code PWR
UDI-Device Identifier00801902101051
UDI-Public(01)00801902101051(11)250527(17)270428(10)33F25B0324(21)
Combination Product (Y/N)N
Initial Reporter CountryAS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberPR-35041-HPHNM
Device Lot Number33F25B0324
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/27/2026
Supplement Date Received by Manufacturer05/06/2026
Initial Report FDA Received Date03/19/2026
Supplement Report FDA Received Date05/12/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/27/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NOT REPORTED
Patient SexUnknown
-
-