| Model Number |
W1TR01 |
| Medical Device Problem Codes |
Pacing Problem (1439); Device Sensing Problem (2917)
|
| Health Effect - Clinical Codes |
Bradycardia (1751); Tachycardia (2095); Syncope/Fainting (4411)
|
| Date of Event |
12/01/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Continuation of d10: 5076-52 lead, implanted: (b)(6) 2013.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the cardiac resynchronization therapy pacemaker (crt-p,) detected ventricular tachycardia (vt) that was suspected to be supra ventricular tachycardia (svt) or dual tachycardia.Review of the electrograms (egm) confirm atrial rates below the lower rate limit (lrl).In addition, the right atrial (ra) lead exhibited undersensing during atrial tachycardia/atrial fibrillation (at/af) events resulting in unnecessary/inappropriate atrial pacing at times.In addition, low p-waves were noted.The crt-p and ra lead remain in use.The patient also experienced syncope with a heart rate of 38bpm.
|
| |
|