| Model Number |
G30474 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/09/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3 - device evaluation anticipated but not yet begun.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
| |
|
Event or Problem Description
|
|
Upon opening the device packaging, it was reported that the distal end of a hiwire nitinol hydrophilic wire guide was observed to be broken prior to use in a ureteral stone procedure.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.Additional information regarding the event has been requested but is currently unavailable.
|
| |
|
Search Alerts/Recalls
|