| Model Number |
VICM5 13.2 (-6.50) |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
01/13/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H11- manufacturer narrative: secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.H6- health effect- clinical code: 4581- significant reduction of iridocorneal angles and significant shallowing of ac.Claim# (b)(4).
|
| |
|
Event or Problem Description
|
|
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced excessive vault, significant reduction of iridocorneal angles and significant shallowing of ac.In a separate surgery the lens was exchanged with a shorter length (vertically implanted) and the problem was resolved.The cause of the event was reported was a patient related factor.
|
| |
|
Search Alerts/Recalls
|