Model Number WA2T412A
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/26/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, it is likely the most probable cause of the malfunction is traced to component failure due to stress.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
It was observed that during the device evaluation, the rigid endoscope telescope exhibited chipped lens in optical system.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TELESCOPE "OES ELITE", 4 MM, 12°, HD, AUTOCLAVABLE
Common Device Name
RIGID ENDOSCOPE TELESCOPE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM   22045
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key24641659
Report Number9610773-2026-02012
Device Sequence Number10207746
Product Code FGC
UDI-Device Identifier04042761077507
UDI-Public04042761077507
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K151191
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 03/19/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberWA2T412A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/03/2025
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/28/2026
Initial Report FDA Received Date03/19/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/30/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1