Model Number 1001.001
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Laceration(s) (1946)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
As the product in question complies with the specification and no deviations were found that could cause or contribute to the reported incident, we suspect that maybe the pinning technique was not optimal, as described in the product's instruction manual (secure the patient's head to the skull clamp): "adjust the skull clamp to the width of the patient's head in the manner that the two skull pins in the rocker arm are equidistant from the centerline of the head and the single skull pin at the extension assembly is in line with this centerline.".
 
Event or Problem Description
Customer informed us on march 17 that one of our products was involved in an incident that resulted in a laceration.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DORO® SKULL CLAMP
Common Device Name
DORO® QR3 SKULL CLAMP
Manufacturer (Section D)
BLACK FOREST MEDICAL GMBH
boetzinger strasse 86
freiburg im breisgau, baden-wuerttemberg 79111
GM  79111
Manufacturer (Section G)
BLACK FOREST MEDICAL GMBH
boetzinger strasse 86
freiburg im breisgau, baden-wuerttemberg 79111
GM   79111
Manufacturer Contact
nicholas preissler
boetzinger strasse 86
freiburg im breisgau, baden-wuerttemberg 79111
GM   79111
MDR Report Key24647273
Report Number3003923584-2026-00011
Device Sequence Number9842593
Product Code HBL
UDI-Device Identifier04250435506196
UDI-Public04250435506196
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K203505
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1001.001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/17/2026
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/17/2026
Initial Report FDA Received Date03/20/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/17/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-