| Model Number |
305U227 |
| Medical Device Problem Codes |
Gradient Increase (1270); Intravalvular regurgitation (4068)
|
| Health Effect - Clinical Codes |
Atrial Fibrillation (1729); Aortic Valve Insufficiency/ Regurgitation (4450)
|
| Date of Event |
03/10/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that an unknown duration post implant of this 27mm aortic bioprosthetic valve, the patient is being evaluated for a transcatheter valve-in-valve replacement.The reason for evaluation was reported as severe aortic central regurgitation, atrial fibrillation with rapid ventricular response (rvr), ejection fraction of 56% and a mean gradient of 9 mmhg.Subsequently on (b)(6) 2026, the patient successfully had a 26mm aortic transcatheter valve-in-valve replacement.It was further noted post procedure, echocardiogram showed patient had no evidence of aortic regurgitation and a mean gradient of 4 mmhg.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Patient's weight and device serial number updated.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|