| Model Number |
M00521421 |
| Medical Device Problem Codes |
Break (1069); Activation, Positioning or Separation Problem (2906)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/25/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a mantis clip was used during an endoscopy procedure on (b)(6) 2026.During the procedure, the mantis clip locked onto tissue, but the handle and spring wire broke, and manipulation of moving left and right, trying to reposition catheter, trying to bend to angle and pull trying to torque to get the clip to release.The clip ultimately was released after a reported 20 minutes of manipulation.The procedure was completed with the same device.There were no patient complications reported as a result of this event.Note: this report pertains to the first of two clips used.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h6: imrdf code a15 captures the reportable event of difficult to release from catheter.Block h11: investigation results.The device was not returned for analysis, as it was disposed; therefore, a technical analysis could not be performed.The complaint device was not returned as it was disposed, therefore, product analysis could not be carried out, for this reason and due to the lack of evidence is unable to confirm the reported event.It was confirmed this device met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of "unable to exclude device problem".
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a mantis clip was used during an endoscopy procedure on (b)(6) 2026.During the procedure, the mantis clip locked onto tissue, but the handle and spring wire broke, and manipulation of moving left and right, trying to reposition catheter, trying to bend to angle and pull trying to torque to get the clip to release.The clip ultimately was released after a reported 20 minutes of manipulation.The procedure was completed with the same device.There were no patient complications reported as a result of this event.Note: this report pertains to the first of two clips used.Additional information, these two clips were used different procedures.
|
| |
|
Additional Manufacturer Narrative
|
|
Block b5: correction updated b5.Block h6: imrdf code a15 captures the reportable event of difficult to release from catheter.Block h11: investigation results the device was not returned for analysis, as it was disposed; therefore, a technical analysis could not be performed.The complaint device was not returned as it was disposed, therefore, product analysis could not be carried out, for this reason and due to the lack of evidence is unable to confirm the reported event.It was confirmed this device met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of "unable to exclude device problem".
|
| |
|
Search Alerts/Recalls
|