• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. MCRYL+ UD 18IN 3-0 S/A PS-2 PRM MP; SUTURE, ABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. MCRYL+ UD 18IN 3-0 S/A PS-2 PRM MP; SUTURE, ABSORBABLE, SYNTHETIC Back to Search Results
Catalog Number MCP497G
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/26/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent an unknown procedure on (b)(6) 2026 and suture was used.During the procedure the needle fell off the suture four different times, like four different strands they pulled.They pulled it out of the case, the appeal pack, and when they want to go arm it, the needle popped off.No x-ray was done but they were able to retrieve the needle.No patient consequences.Device is returning.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 3/20/2026.H6 component code: g07002 - pending evaluation of returned device.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: did the needles fall into the patient? if yes, was the needle piece(s) retrieved during the same procedure? were x-rays taken to locate the needle piece(s)? what measures were taken to retrieve the broken piece? was there any additional tissue damage as a result of searching for the needle piece? please provide the source or name and title of external person providing answers to follow-up (external person submitting answers to sales re to date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 4/20/2026.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H6 component code: g07002 no device problem found.Additional information: d9, h3, h6.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.H3 investigational summary: the product was returned to ethicon for evaluation.One open box with nine representative unopened samples were received for analysis.Product code mcp497g.The complaint sample was not received for evaluation.Upon initial inspection of the samples, no external damages were observed on the external packets.In order to evaluate the conditions of the returned samples, the packets were opened, and no defects were detected.The swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along the strand to detect any issue related no defects were observed during evaluation.A functional test was performed using instron equipment, and the pull force results were above the minimum requirements.The event described could not be confirmed as the device performed without any defect noted.It should be noted that as part of our quality process, each batch is randomly inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.The event described could not be confirmed as no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MCRYL+ UD 18IN 3-0 S/A PS-2 PRM MP
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
calle durango #2751
ciudad juarez 32575
MX   32575
Manufacturer Contact
orla o'mahony
1000 route 202
raritan, NJ 08869
3035526892
MDR Report Key24648706
Report Number2210968-2026-02848
Device Sequence Number13235413
Product Code GAM
UDI-Device Identifier10705031046603
UDI-Public(01)10705031046603(17)270630(10)109UEH
Combination Product (Y/N)Y
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K050845
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberMCP497G
Device Lot Number109UEH
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/19/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/25/2026
Supplement Date Received by Manufacturer03/29/2026
Initial Report FDA Received Date03/20/2026
Supplement Report FDA Received Date04/20/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/23/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-