| Catalog Number |
00598604702 |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Pain (1994); Inadequate Osseointegration (2646); Aseptic loosening (4883)
|
| Date of Event |
04/20/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).D10: unknown - unknown femoral component - unknown.Unknown - unknown articular surface - unknown.Attempts have been made to gather all product identification information and no further information has been provided.The event cannot be confirmed.No product was returned, or pictures provided; visual and dimensional evaluations could not be performed.The device history record was unable to be performed as the lot number of the device involved in the event is unknown.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent a left knee arthroplasty.Approximately 3 years post-op, the patient is alleging experiencing pain and loosening and is planning to be revised.No revision has been reported at this time.Attempts have been made and all available information has been provided.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient had an initial left total knee arthroplasty performed.Approximately 4 years post-op, the patient was revised due to mechanical failure.During the revision, the tibial component was found to be grossly loose with no cement adherence to the implant.The femoral and patellar components were well fixed.The tibial component and articular surface were revised without complications.Attempts have been made and all available information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated:
a2; b4; b5; b7; d1; d2; d4; d6a; d10; e1; e2; e3; g1; g3; g4; g6; h1; h2.D10:
00596401751 - femoral component option size g left compatible with lps-flex prolong - 64028394.00596404212 - articular surface size gh 12 mm height''"use with plate 5 - 64507192.00597206535 - all poly petella standard cemented size 35 mm diameter 9.0 mm thickness - 65270040.Unknown - unknown cobalt cement - unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated:
a1; a3; b3; b4; b5; b7; d2; d6b; e1; g1; g3; g6; h1; h2; h6.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|