| Model Number |
42759 |
| Medical Device Problem Code |
Premature Separation (4045)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/26/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that while attempting to advance the icast stents, the stent dislodged from the balloon.Two stents were successfully removed from the patient after dislodging and traveling into the sheath.The third stent was implanted in the patient.There was no patient harm reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Due to character restrictions in following section.D4 field: additional two stent involved in the event with model no's 42959, lot no's 521342.It is unknown currently which of the stent is been implanted.E1: event site name: (b)(6) hospital.A supplemental report will be submitted upon completion of our investigation.
|
| |
|
Event or Problem Description
|
|
Additional information states that the stent was successfully implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Due to character restrictions in following section.D10: 8fr terumo destination, 18fr medtronic aptus heli-fx sheath.Related mdrs: 3011175548-2026-0000027, 3011175548-2026-0000028.Updated fields: a2, b5, d6a, h6 (type of investigation).
|
| |
|