| Model Number |
9736428 |
| Medical Device Problem Code |
Device Sensing Problem (2917)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/12/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Medtronic received information regarding a navigation system being used during a sacroiliac and thoracolumbar procedure.It was reported that trajectory 1 and 2 intermittently showed blank/black.The reference frame "jumped around" on tracking detail.The manufacturer representative (rep) confirmed that instruments not being used were out of view, that there were fingerprints on the right camera lens, that one sphere on the patient reference frame was red, and that the system was connected to a boom.The rep moved the camera in and out, pressed spheres down without using excessive force as the spheres were indirectly connected to the patient's spine, cleaned the camera lenses, and cleaned instruments.The issue did not replicate with another patient reference frame and no apparent issues were seen in the network device interface (ndi) toolbox.There was no surgical delay.There was no impact on patient outcome.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: the instrument (product #: 9736428) has not been returned to the manufacturer for evaluation.Codes b17, c20, and d15 are applicable.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|