PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH SPO2 CABLE; OXIMETER Back to Search Results
Model Number SP02 8-PIN D-SUB ADAPTER CABLE 3M (8PIN)
Medical Device Problem Code Low Readings (2460)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/16/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The customer returned three cables of the same lot code 0625 to the philips failure analysis lab for evaluation.The evaluation included visual inspection, electrical testing, and functional performance testing to reproduce the reported issue.Visual inspection found that the cable has no visible physical damage.Functional testing confirmed that the cable could not produce a steady signal.The voltage test on pins from the 9-pin connector pin 6 to the 8-pin connector pin 1 showed no voltage reading using a hp e2377a multimeter.Further functional testing confirmed the voltage was outside the tolerance.Based on these findings, the adapter cable did not perform as designed, and the customer¿s complaint was confirmed.The affected adapter cable will be scrapped at the mcs failure analysis laboratory, as no further testing is required.Based on the information available and results of additional analysis, no further action is necessary at this time.For reports on the other two cables returned, please refer to manufacturer's references: (b)(4).
 
Event or Problem Description
Philips received a complaint on an sp02 8-pin d-sub adapter cable 3m (8pin) indicating that the spo2 cord will not work.No matter the patient or position, the cord relays a weak pleth signal or no signal and the o2 is always reported low.The device was in use on a patient at time of event, there was no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPO2 CABLE
Common Device Name
OXIMETER
Manufacturer (Section D)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM  71034
Manufacturer (Section G)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
Manufacturer Contact
hauke schik
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
70314632031
MDR Report Key24650072
Report Number9610816-2026-100549
Device Sequence Number9842685
Product Code DQA
UDI-Device Identifier00884838011946
UDI-Public(01)00884838011946(10)0625
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K032979
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSP02 8-PIN D-SUB ADAPTER CABLE 3M (8PIN)
Device Catalogue Number989803128651
Device Lot Number0625
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/13/2026
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/19/2026
Initial Report FDA Received Date03/20/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/01/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-