| Model Number |
7731 |
| Medical Device Problem Code |
High impedance (1291)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/04/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this pacemaker system triggered a lead safety switch (lss) due to high out of range right ventricular (rv) lead pacing impedance measurement greater than 3000 ohms.Technical services (ts) was consulted and recommended further clinical evaluation.This rv lead remains in service.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis; therefore, a technical analysis could not be performed.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of known inherent risk.This investigation will be updated should pertinent information become available in the future.
|
| |
|