Model Number QVM2-D-HM
Medical Device Problem Code Incorrect Measurement (1383)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/23/2026
Type of Reportable Event Malfunction
Event or Problem Description
User reported an error with the pressurement measurement on the ventilation module.The value from the pressure sensor did not match an external pressure sensor.There was no patient harm.
 
Additional Manufacturer Narrative
The investigation determined that the settings were configured to use the ventilation module with the spectrum medical oxygenator, meaning both the sweep and po2 ports are used on the quantum ventilation module.The customer's default settings are intended to be used with a different oxygenator.Therefore, there was no device malfunction, but the settings were corrected and the device was confirmed to operate as expected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUANTUM VENTILATION MODULE
Common Device Name
GAS CONTROL UNIT, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
SPECTRUM MEDICAL LTD
ashville business park
commerce road
gloucester, gloucestershire GL2 9 SL
UK  GL2 9SL
Manufacturer (Section G)
SPECTRUM MEDICAL LTD
ashville business park
commerce road
gloucester, gloucestershire GL2 9 SL
UK   GL2 9SL
Manufacturer Contact
cameron schmitt
481 munn road
suite 180
fort mill, SC 29715
MDR Report Key24650378
Report Number3006073153-2026-00092
Device Sequence Number21372381
Product Code DTX
UDI-Device Identifier05060434422071
UDI-Public05060434422071
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K202733
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Remedial Action Repair
Type of Report Initial
Report Date (Section B) 03/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberQVM2-D-HM
Device Catalogue Number51-000082-00
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/23/2026
Initial Report FDA Received Date03/20/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/12/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-