| Catalog Number |
11-107018 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Joint Dislocation (2374)
|
| Date of Event |
02/26/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).G2: foreign - event occurred in australia.H6: proposed component code: mechanical (g04) ¿ head.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that approximately three months post-implantation, the patient underwent a hip revision due to a dislocation.The head and liner were revised.Attempts have been made and no further information has been provided.
|
| |
|
Event or Problem Description
|
|
No further information is available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: h2; h3; h6; h10.Complaint is not confirmed.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record identified no deviations or anomalies during manufacturing.Review of complaint history identified additional similar complaints for the reported items and no additional complaints for the reported part and lot combinations.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|