| Catalog Number |
03-2722-9C |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Hemorrhage/Blood Loss/Bleeding (1888)
|
| Date of Event |
03/06/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
|
| |
|
Event or Problem Description
|
|
A user facility¿s registered nurse (rn) reported that a fresenius combiset bloodline had a loose connection between the clic blood chamber window and the red arterial port connector piece resulting in blood loss during a patient¿s hemodialysis (hd) treatment.The issue occurred immediately after the initiation of treatment.The machine, a fresenius 2008t machine, did not alarm but was not expected to.A fresenius dialyzer was also in use.There were no loose connections or issues during priming.There were no changes to pressure or blood flow rate during treatment.There was no visible damage to the bloodline.The patient¿s estimated blood loss (ebl) was approximately 5ml.The patient was restarted on the same machine and treatment completed successfully with new supplies.There was no patient serious injury or medical intervention required as a result of this event.The complaint device is not available to be returned to the manufacturer for evaluation.
|
| |
|
Event or Problem Description
|
|
A user facility¿s registered nurse (rn) reported that a fresenius combiset bloodline had a loose connection between the clic blood chamber window and the red arterial port connector piece resulting in blood loss during a patient¿s hemodialysis (hd) treatment.The issue occurred immediately after the initiation of treatment.The machine, a fresenius 2008t machine, did not alarm but was not expected to.A fresenius dialyzer was also in use.There were no loose connections or issues during priming.There were no changes to pressure or blood flow rate during treatment.There was no visible damage to the bloodline.The patient¿s estimated blood loss (ebl) was approximately 5ml.The patient was restarted on the same machine and treatment completed successfully with new supplies.There was no patient serious injury or medical intervention required as a result of this event.The complaint device is not available to be returned to the manufacturer for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
Plant investigation: as the device was not returned to the manufacturer, a physical evaluation could not be performed.A batch records review was conducted by the manufacturer for the reported lot.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.The entire lot has been sold and distributed.In addition, a device history review was performed and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The lot met all specifications for release.A product history review did not reveal a probable cause for the customer complaint.As a physical evaluation could not be performed, a definitive conclusion regarding the reported incident could not be reached and a cause could not be confirmed.
|
| |
|