• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PTS DIAGNOSTICS CARDIOCHEK PLUS ANALYZER; CLINICAL CHEMISTRY ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PTS DIAGNOSTICS CARDIOCHEK PLUS ANALYZER; CLINICAL CHEMISTRY ANALYZER Back to Search Results
Model Number 1.11
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/18/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
During the investigation, it is determined that the terminal displacement is most likely caused by the customer providing undue force onto the battery terminals.In addition, the potential exists for the battery contact to lose traction with the compartment causing less force to dislodge however improper removal of batteries will still force the dislodge.Several key factors point to this decision: 1.The battery terminals would have to have been properly seated at the time of manufacturing because the analyzer would not power on and would consequently be unable to be calibrated and released.2.Upon close inspection of the battery terminal area, it was observed that the terminals were fully seated initially due to visible drag marks created by the metal terminal scraping the plastic casing when excessive force was applied.These markings are present from the point of terminal seating down to where the terminals lie currently.Pts is confident that this damage occurred post-distribution, as there are no manufacturing steps that would cause this type of damage.
 
Event or Problem Description
There were no allegations of patient harm or incorrect results.The customer reported that their cardiochek plus analyzer "keeps getting warm to touch".The reported analyzer was being used beyond it's recommended device life at the time the incident occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOCHEK PLUS ANALYZER
Common Device Name
CLINICAL CHEMISTRY ANALYZER
Manufacturer (Section D)
PTS DIAGNOSTICS
4600 anson blvd
whitestown
Manufacturer (Section G)
PTS DIAGNOSTICS
4600 anson blvd
whitestown IN 46075
Manufacturer Contact
randall cundiff
4600 anson blvd
whitestown, IN 46075
MDR Report Key24650626
Report Number1836135-2026-00003
Device Sequence Number21272398
Product Code CGA
UDI-Device Identifier00381932700016
UDI-Public(01)00381932700016(8012)1.11
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
K140068
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1.11
Device Catalogue Number2700
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/02/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/18/2026
Initial Report FDA Received Date03/20/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-