BAXTER HEALTHCARE CORPORATION VOLARA SYSTEM; DEVICE, POSITIVE PRESSURE BREATHING, INTERMITTENT
|
Back to Search Results |
|
| Model Number |
M08594A |
| Medical Device Problem Code |
Crack (1135)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/25/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The baxter technical support found the circuit and filter needed to be replaced.The volara¿ system, expands the lungs and moves mucus from the airways.Proven1 in the hospital, it¿s highly effective airway clearance therapy that takes just 10 minutes to deliver.With the volara¿ system, patients with chronic conditions can get the individualized treatment they need at home, with physician-prescribed presets that make therapy easy to follow.A search of the baxter maintenance records did not show baxter performed any preventative maintenance on this product.It is unknown if the customer performs preventative maintenance on their products.A replacement patient circuit and filter was sent to the customer.Based on this information, no further actions are required at this time.Although there was no reported injury with this event, if the report of a broken filter were to recur, it could potentially cause serious injury or death.Therefore baxter is reporting this event.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2026, a patient contacted technical service to report that volara system serviced, hc,na (product code m08594a, serial number (b)(6)), had smart filter cracked.This occurred at home during patient use.There was no patient injury or death reported with this event.
|
| |
|