• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS INTELLIVUE MX40 WLAN; MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS INTELLIVUE MX40 WLAN; MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS) Back to Search Results
Model Number MX40 - WLAN
Medical Device Problem Code No Audible Alarm (1019)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/25/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A philips bench repair technician (brt) evaluated the device and confirmed that there was no speaker sound at start up test and the device failed at manual power on test.The brt replaced the speaker assembly to resolve the issue.The device was operational after repairs were completed and the device was returned to the customer.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Event or Problem Description
Philips received a complaint on the mx40 1.4 ghz smart hopping indicating there was no audio.A good faith effort (gfe) conducted confirmed that the device was completely out of sound.The device was not in use at the time of the event.No adverse event occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIVUE MX40 WLAN
Common Device Name
MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer Contact
deborah currlin
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key24651133
Report Number1218950-2026-100159
Device Sequence Number15913298
Product Code MHX
UDI-Device Identifier00884838030350
UDI-Public(01)00884838030350
Combination Product (Y/N)N
Initial Reporter StateID
Initial Reporter CountryUS
PMA/510(K) Number
K113125
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 03/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMX40 - WLAN
Device Catalogue Number865352
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/10/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/25/2026
Initial Report FDA Received Date03/20/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/22/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-