• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH DEPTH GAUGE FOR SCREWS Ø 2.7 TO 4.0 MM, MEASURING RANGE UP TO 60 MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH DEPTH GAUGE FOR SCREWS Ø 2.7 TO 4.0 MM, MEASURING RANGE UP TO 60 MM Back to Search Results
Catalog Number 319.01
Medical Device Problem Codes Incorrect Measurement (1383); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.D4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.
 
Event or Problem Description
It was reported on an unknown date that several depth gages were bent/no longer accurate.The va guides were wearing down and not staying in va holes during drilling without assistance.The screwdriver shafts were worn down.The usf handles were spinning on the screw when the surgeon was tightening down screws.The trocars were bent/unstable.There was no surgical delay.Action was taken by removing the items from service, same instruments from other trays were used instead.The procedure was successfully completed.There were no fragments generated with no patient involvement.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: d10 concomitant.H6 component codes: most relevant component code is g07002 (appropriate term/code not available) to capture no findings available due to no product returned.Investigation summary: product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEPTH GAUGE FOR SCREWS Ø 2.7 TO 4.0 MM, MEASURING RANGE UP TO 60 MM
Common Device Name
DEPTH GAUGE
Manufacturer (Section D)
SYNTHES GMBH
luzernstrasse 19-21
zuchwil 4528
SZ  4528
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
elba bello
1302 wright lane east
west chester, PA 19380
9083863534
MDR Report Key24651900
Report Number8030965-2026-02707
Device Sequence Number10168565
Product Code HTJ
UDI-Device Identifier10886982189950
UDI-Public(01)10886982189950
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number319.01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/04/2026
Supplement Date Received by Manufacturer04/09/2026
Initial Report FDA Received Date03/20/2026
Supplement Report FDA Received Date04/11/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DEPTH GAUGE FOR LARGE SCREWS,; DEPTH GAUGE FOR LARGE SCREWS,; DEPTH GAUGE FOR LARGE SCREWS,; DEPTH GAUGE FOR SCREWS Ø 2.7 TO 4.0 MM,; DEPTH GAUGE FOR SCREWS Ø 2.7 TO 4.0,; HANDLE W/QUICK-COUPL,; HANDLE W/QUICK-COUPL,; HANDLE W/QUICK-COUPL.; SCRDRIVER SHAFT SELF-HOLDING L165 F/STAR,
Patient SexUnknown
-
-