| Model Number |
M00558700 |
| Medical Device Problem Code |
Burst Container or Vessel (1074)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/17/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a cre pro wireguided dilatation balloon was attempted to be used during a dilatation procedure performed on (b)(6) 2026.During the procedure, it was reported that the balloon burst.The procedure was completed with an alternative device.No further information has been obtained despite good faith efforts.There were no reported patient complications as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h6: imdrf device code a0402 captures the reportable event of a balloon burst.Block h11: investigation results: the device was not returned for analysis and was reported to be disposed; therefore, a technical analysis could not be performed.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.Review of the instructions for use (ifu) confirmed there was relevant content and sufficient guidance with respect to the circumstances described within this complaint.No updates are required to the ifu as a result of this event.A risk review was completed and confirmed that the event of " balloon burst" was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of unable to exclude device problem.
|
| |
|
Search Alerts/Recalls
|