DEPUY MITEK LLC US MINILOK QA+ # 2-0 OC V5; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE Back to Search Results
Catalog Number 212853
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/06/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an arthroscopic ligament repair procedure, it was discovered that the anchor on the minilok qa+ # 2-0 oc v5 device broke off.All the broken parts were removed.There were no delays in the surgical procedure.It was reported that a spare device was used to complete the surgery.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11: additional narrative: e1: the reporter¿s complete facility address was not provided.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11: additional narrative: d9, h3, h6: the actual device has been returned and is currently pending evaluation.Once the investigation has been completed, and if additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11: additional narrative: investigation summary: the product was returned to depuy synthes for evaluation.Visual inspection found that the device had the anchor and sutures detached from the inserter.No damage or defects were found in the anchor.The shaft and handle did not show anomalies.The device had foreign matter, presumably biological matter.No other anomalies were noted.The overall complaint was not confirmed as the observed condition of minilok qa+ # 2-0 oc v5 would have not contributed to the complained device issue.Based on the findings, the investigation could not draw any conclusions and no potential cause about the reported event can be established due to the insufficient evidence provided.It has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review: there was no non-conformance regarding this lot.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINILOK QA+ # 2-0 OC V5
Common Device Name
FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
elba bello
325 paramount drive
raynham, MA 02767
9083863534
MDR Report Key24652383
Report Number1221934-2026-01187
Device Sequence Number20265537
Product Code MAI
UDI-Device Identifier10886705002429
UDI-Public(01)10886705002429(10)109QQC(17)280630
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number212853
Device Lot Number109QQC
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/13/2026
Supplement Date Received by Manufacturer04/06/2026
04/10/2026
Initial Report FDA Received Date03/20/2026
Supplement Report FDA Received Date04/08/2026
04/13/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/17/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-