| Catalog Number |
TSX6B8 |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/05/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the implant at tooth site #14 loosened.The product was a screw retained restoration on the same implant, which is being returned to aid in the investigation.As this case had been inherited from another dental office, no additional information was provided concerning the previous restorations or about the crown or abutment details.The patient will return for an implant replacement.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimvie complaint number: (b)(4).A4: weight unknown / not provided.Customer has indicated that the product is in process of being returned to zimvie for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted to relay additional information and device evaluation.The following sections are being reported: b4: date of this report was updated.D9: device availability and return date were updated.G3: date received by manufacturer was updated.G6: type of report was updated.H2: type of follow up was updated.H3: device evaluated by manufacturer was updated.H6: type of investigation code was added: 10, 4109, 3331 and 4111.H6: investigation findings code was added: 213.H6: investigation conclusions code was added: 4315.H11: narrative/data was updated.Zimvie received the reported device for evaluation.Visual evaluation and functional testing was performed , no damaged threads has been identified.Dhr review was completed for the subject lot number.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number for similar events and no other complaint was identified.The customer did not submit images for the reported event.A definitive root cause could not be determined since the device malfunction did not occur, and the reported event has been unconfirmed following functional testing and physical evaluation.Therefore, based on the available information, a device malfunction did not occur.There was no damage identified to the threads.The reported event was not confirmed.No further investigation or immediate capa/hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.
|
| |
|
Event or Problem Description
|
|
No additional event information received at the time of this report.
|
| |
|
Search Alerts/Recalls
|