MAUDE Adverse Event Report: ZIMVIE US CORP LLC TSX IMPLANT, 6.0MMD, 8MML; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMVIE US CORP LLC TSX IMPLANT, 6.0MMD, 8MML; DENTAL IMPLANT Back to Search Results
Catalog Number TSX6B8
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/05/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the implant at tooth site #14 loosened.The product was a screw retained restoration on the same implant, which is being returned to aid in the investigation.As this case had been inherited from another dental office, no additional information was provided concerning the previous restorations or about the crown or abutment details.The patient will return for an implant replacement.
 
Additional Manufacturer Narrative
Zimvie complaint number: (b)(4).A4: weight unknown / not provided.Customer has indicated that the product is in process of being returned to zimvie for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Additional Manufacturer Narrative
This report is being submitted to relay additional information and device evaluation.The following sections are being reported: b4: date of this report was updated.D9: device availability and return date were updated.G3: date received by manufacturer was updated.G6: type of report was updated.H2: type of follow up was updated.H3: device evaluated by manufacturer was updated.H6: type of investigation code was added: 10, 4109, 3331 and 4111.H6: investigation findings code was added: 213.H6: investigation conclusions code was added: 4315.H11: narrative/data was updated.Zimvie received the reported device for evaluation.Visual evaluation and functional testing was performed , no damaged threads has been identified.Dhr review was completed for the subject lot number.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number for similar events and no other complaint was identified.The customer did not submit images for the reported event.A definitive root cause could not be determined since the device malfunction did not occur, and the reported event has been unconfirmed following functional testing and physical evaluation.Therefore, based on the available information, a device malfunction did not occur.There was no damage identified to the threads.The reported event was not confirmed.No further investigation or immediate capa/hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.
 
Event or Problem Description
No additional event information received at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TSX IMPLANT, 6.0MMD, 8MML
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMVIE US CORP LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMVIE US CORP LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
mark mashburn
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key24653093
Report Number0001038806-2026-01509
Device Sequence Number6109453
Product Code DZE
UDI-Device Identifier00844868045328
UDI-Public(01)00844868045328(17)250524(10)1267798
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K220978
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other
Device Expiration Date05/24/2025
Device Catalogue NumberTSX6B8
Device Lot Number1267798
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date03/20/2026
Supplement Report FDA Received Date04/14/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/25/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
Patient SexFemale
Patient EthnicityNon Hispanic
-
-