Model Number TSGC0801
Medical Device Problem Code Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/24/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to the reported difficult to remove.Additionally, a review of the complaint history identified similar complaints reported to this lot, and this device appears to be related to a potential product quality issue.The investigation determined that the reported difficult to remove may be related to a potential product quality issue.This complaint falls within the scope of an exception, as the complaint description and device code match the specific issue described in the exception.The issue is being addressed per internal operating procedures.Abbott will continue to trend the performance of these devices.
 
Event or Problem Description
It was reported that on (b)(6) 2026, a patient presented with grade 4-5 functional tricuspid regurgitation (tr) for a triclip procedure.After completing the procedure, it was unable to remove the triclip steerable guide catheter (tsgc) from the stabilizer.Tsgc was removed from the patient when it was still attached to the stabilizer.After removing from the patient, it needed more force to pull tsgc from the stabilizer.All steps were followed as per ifu.The final tr was grade 1-2.There were no adverse patient effects or clinically significant delays reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRICLIP¿ G5 STEERABLE GUIDE CATHETER
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key24653426
Report Number2135147-2026-01801
Device Sequence Number17574441
Product Code DQY
UDI-Device Identifier05415067050724
UDI-Public(01)05415067050724(17)270430(10)51031R1052
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K243224
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTSGC0801
Device Catalogue NumberTSGC0801
Device Lot Number51031R1052
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/24/2026
Initial Report FDA Received Date03/20/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/31/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
Patient Weight80 KG
-
-