| Model Number |
M00584170 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/23/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a trutome 44 was used in the ampulla during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2026.During the procedure, the cutting wire of the trutome 44 broke.It was reported that no part of the cutting wire detached and fell into the patient.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h6: imdrf device code a0401 captures the reportable event of cutting wire broken.Block h11: investigation results the device was not returned for analysis and was reported to be disposed; therefore, a technical analysis could not be performed.A review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of unable to exclude device problem.
|
| |
|