| Catalog Number |
31251180S |
| Medical Device Problem Code |
Positioning Failure (1158)
|
| Health Effect - Clinical Codes |
Pain (1994); Perforation (2001)
|
| Date of Event |
11/16/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
As reported by the customer: "when inserting the short titanium gamma nail, it was impossible to lock it in place statically due to friction on the nail; therefore, it had to be locked dynamically, resulting in a high risk of the distal screw breaking when the patient put weight on her leg.On the same nail, the slide for inserting the cephalic screw locking screw became misaligned leading to slippage when aiming.The screw ended up in the patient's buttock and could not be removed.A second prosthesis was opened and presented the same problem." update (b)(6) 2026 was the procedure completed successfully? yes, but the nail was stabilized distally in a dynamic position.The patient reports postoperative pain at the screw site.Was the procedure delayed? yes: 10 minutes of additional operating time.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information has been requested and if received, will be provided in the supplemental report.
|
| |
|
Event or Problem Description
|
|
As reported by the customer: "when inserting the short titanium gamma nail, it was impossible to lock it in place statically due to friction on the nail; therefore, it had to be locked dynamically, resulting in a high risk of the distal screw breaking when the patient put weight on her leg.On the same nail, the slide for inserting the cephalic screw locking screw became misaligned leading to slippage when aiming.The screw ended up in the patient's buttock and could not be removed.A second prosthesis was opened and presented the same problem."
update february 26, 2026
was the procedure completed successfully? yes, but the nail was stabilized distally in a dynamic position.The patient reports postoperative pain at the screw site.Was the procedure delayed? yes: 10 minutes of additional operating time.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could not be confirmed since the device was not returned for evaluation and no other evidences were provided.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If the device is returned or any further information is provided, the investigation report will be reassessed.
|
| |
|