The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd drive pcb fluid damage.Based on the result, we concluded that it was caused due to the fluid damage from the ccd drive pcb.In addition, our technician confirmed that the ultrasound connector body fluid damage, the lg connector fluid damage, the control body corroded, the mechanical base plate corroded, the lg connector corroded, the suction channel kink, the down pulley wire rupture, the up pulley wire rupture, the left pulley wire rupture, the right pulley wire rupture, the ccd module with drive pcb distorted image, and the us probe eus image no scanline; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report "hr-rpt-0586(image failure)" and/or the risk analysis results, it was evaluated to submit mdr.
|