• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY LLC SCHICK33 S2 SENSOR SHIP KIT/3.0 CABLE 6'; UNIT, X-RAY, INTRAORAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY LLC SCHICK33 S2 SENSOR SHIP KIT/3.0 CABLE 6'; UNIT, X-RAY, INTRAORAL Back to Search Results
Catalog Number 100008624
Medical Device Problem Code Poor Quality Image (1408)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
While no serious injury resulted in this event, if this malfunction recurred, it could cause or contribute to a serious injury or require medical or surgical intervention to preclude such.This event, therefore, is reportable per 21cfr part 803.Usb 3 firmware v1.4 contains an issue in the portion of code used to optimize sensor performance.If the optimization occurs during an exposure or sensor readout, there is a chance that the resulting image will be corrupted.Image data following the point of corruption will consist of data that was previously held in the memory of the usb device.As a result, the presented image may contain a portion of data from a previous exposure.In general, this would be obvious to the user since the presented image would either contain obvious artifacts or different anatomy than expected.
 
Event or Problem Description
While the customer was using a part of an intraoral sensor system, they allege there was an x-ray image transmission issue where the x-ray image was not captured after radiation has occurred (transmission failure).No injury was reported from the alleged event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCHICK33 S2 SENSOR SHIP KIT/3.0 CABLE 6'
Common Device Name
UNIT, X-RAY, INTRAORAL
Manufacturer (Section D)
DENTSPLY LLC
1301 smile way
york PA 17404
Manufacturer (Section G)
DENTSPLY LLC
1301 smile way
york PA 17404
Manufacturer Contact
dan eagar
221 west philadelphia st.
york, PA 17401
7178457511
MDR Report Key24660951
Report Number2424472-2026-00610
Device Sequence Number9226547
Product Code EAP
UDI-Device IdentifierD0031000086240
UDI-PublicD0031000086240
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K093453
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 03/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number100008624
Was Device Available for Evaluation? No
Distributor Facility Aware Date03/09/2026
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/09/2026
Initial Report FDA Received Date03/23/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-