• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER LEIBINGER FREIBURG FACIAL ID-ORTHOG-2PLATES; PLATE, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER LEIBINGER FREIBURG FACIAL ID-ORTHOG-2PLATES; PLATE, BONE Back to Search Results
Catalog Number 78-92002
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/23/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the surgeon removed the broken hardware and placed new hardware.
 
Additional Manufacturer Narrative
Device is not available for evaluation.If additional information is received, it will be reported on a supplemental report.
 
Event or Problem Description
No new information.
 
Additional Manufacturer Narrative
Additional information: h4, h6 correction: d2a the reported plate fracture could not be confirmed because the devices were not returned and no postoperative imaging was available for review.The investigation found no evidence of a manufacturing or design related issue.Available information indicates the plates met all design and manufacturing requirements, and the implants had been in situ for more than one year before revision surgery.Based on the statistical investigation, there is no indication for an incorrectly working product or any design, material, or manufacturing related issue.Therefore, no corrective and/or preventive actions are deemed necessary at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FACIAL ID-ORTHOG-2PLATES
Common Device Name
PLATE, BONE
Manufacturer (Section D)
STRYKER LEIBINGER FREIBURG
boetzingerstr. 41
freiburg D-791 11
GM  D-79111
Manufacturer (Section G)
Stryker Leibinger Freiburg
boetzingerstr. 41
freiburg D-791 11
GM   D-79111
Manufacturer Contact
elena gehring
boetzingerstr. 41
freiburg D-791-11
GM   D-79111
+49()7614888208
MDR Report Key24661192
Report Number0008010177-2026-00038
Device Sequence Number9344462
Product Code JEY
UDI-Device Identifier07613327370935
UDI-Public07613327370935
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial,Followup
Report Date (Section B) 09/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Catalogue Number78-92002
Device Lot Number2411191102
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/23/2026
Supplement Date Received by Manufacturer02/23/2026
Initial Report FDA Received Date03/23/2026
Supplement Report FDA Received Date09/16/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/03/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-