| Model Number |
E2450H-GNSB |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/24/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D10 concomitant product: l1818-04p01cn-1, lap sponge; xray; rfd; 18x18; nstrl; lf; (lot #250911at).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during a procedure, a fire occurred in one of the operating rooms, and the surgeon wondered if it could have been caused by the bovie touching an x-ray detectable lap sponge.The bovie started the fire (no flame) that produced a 2 in round hole in the lap sponge.The sponge was tossed onto the floor.In the general pack, a bovie pencil was used along with an x-ray detectable lap sponge.There was no patient injury.
|
| |
|