|
There has been a previous report received where lack of water flow has caused an overheating insert.Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.Investigation results return qty (b)(4), 25073, 30k fsi-sli-1000 fg-63, 00136238 bul per mfg.Date.Test method / gp005-wi02 inductance: gauge id# 5349 due: 12/31/2026 result 3.15 ¿ meets spec ¿ does not meet spec ¿ n/a tip condition: ¿ meets spec ¿ does not meet spec ¿ n/a stack condition ¿ meets spec ¿ does not meet spec ¿ n/a tested on: g130 gage id # (b)(4) due date: (b)(6) 2026 set pressure: 35 psi water flow: output: 35 psi - ¿ meets spec ¿ does not meet spec ¿ n/a spray: - ¿ meets spec ¿does not meet spec ¿ n/a water leak: ¿ hp sealing ring ¿ proximal ring ¿ distal ring ¿ seam leak ¿ n/a vibration: ¿ meets spec ¿ does not meet spec ¿ n/a grip condition: ¿ meet spec.¿ does not meet spec ¿ n/a other visual observation: insert tip is bent.Insert water temperature was tested on a digital thermometer i.D.# (b)(4).Due date: (b)(6) 2026.The temperature was 85.2 °f, no fault found with temperature.The specifications state the temperature is not to exceed 118.4°f.(per pds-8016 (rev 6).Insert tip temperature was tested on a digital thermometer i.D.# (b)(4).Due date: (b)(6) 2026.The temperature was 90.2°f, no fault found with temperature.The specifications state the temperature is not to exceed 118.4°f.(per pds-8016 (rev 6).Alleged event: confirmed not confirmed.
|