|
ABBOTT VASCULAR INC. MINI TREK¿; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
|
Back to Search Results |
|
| Model Number |
1012270-12 |
| Medical Device Problem Codes |
Material Rupture (1546); Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/03/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.Production record and corrective and preventative actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a query of the complaint handling database for the reported lot revealed there is no indication of a lot specific issue.Based on the information received, the investigation determined that the reported issue appears to be related to operational context.There was no leak noted during preparation, which suggests a product quality issue did not contribute to the reported difficulties.It was reported by the account that the balloon dilatation catheter (bdc) interacted with the moderately tortuous, moderately calcified and 90% stenosed anatomy resulting in the reported difficulty advancing the device and subsequently the balloon became damaged and ruptured during inflation.Based on the results of the complaint investigation, there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure performed was to treat a de novo, moderately calcified, moderately tortuous left anterior descending coronary artery that is 90% stenosed.During advancement of the 2.00x12mm mini trek rx balloon dilatation catheter (bdc), resistance was encountered with the lesion.During inflation, the balloon ruptured at 6 atmospheres during the first inflation.The mini trek rx bdc was removed and a non-abbott balloon was used to complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Search Alerts/Recalls
|
|
|