| Model Number |
6662 |
| Medical Device Problem Codes |
Premature Discharge of Battery (1057); Battery Problem (2885)
|
| Health Effect - Clinical Codes |
Inadequate Pain Relief (2388); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/26/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that ipg was reaching end of life.Additionally, impedance check revealed high impedances on the right side.As such, surgical intervention took place wherein it was noted that the extensions were completely tangled due to ipg flipping.In turn, the ipg and extensions were explanted and replaced resolving the issues.Reportedly, therapy was confirmed post op.The device was discarded.
|
| |
|
Additional Manufacturer Narrative
|
|
Date of event is estimated.
|
| |
|
Event or Problem Description
|
|
Additional information received indicates that the ipg had reached end of life.
|
| |
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|