|
The reported event could be confirmed with the provided x-ray images in which we can see that the nail is slightly bent and one of the distal screws is broken.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required currently.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design-related problems were found during the investigation.Hcp opinion on provided x rays:
"the event appears consistent with a nonunion scenario associated with inadequate fracture-site stability.The hcp indicated that the fracture pattern likely contributed to reduced construct stability, resulting in continuous motion at the fracture site.This persistent instability may have contributed to screw breakage and subsequent increased loading on the remaining fixation components, potentially leading to nail bending.Overall, the event appears multifactorial, with contributing factors including fracture characteristics, biological factors, surgical technique, and postoperative loading/activity."
based on the limited information available, the root cause of the reported event cannot be defined as this is often muti-factorial: the location of the fracture and the technical aspects of the surgical procedure could have contributed to the event.Furthermore, patient condition and activity levels post-op may also have contributed to an inadequate load distribution, and therefore the breakage of the implant.If the device is returned or if any additional information is provided, the investigation will be reassessed.
|