| Model Number |
NT821731C |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/11/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Initial report ref to natus complaint # (b)(4).(b)(4) rev 13 risk analysis spreadsheet - eds 3 external drainage system hazard 4.9 - the stopcock at the bottom of the burette tube fails during the operation (unable to turn the stopcock valve and/or breakage of the valve) effect (harm): over drainage/under drainage of csf residual risk: medium the hazards identified have been reduced as far as possible, and the benefit of use of the entire product outweighs the risks identified.Related to capa005781.Further investigation to be carried out.
|
| |
|
Event or Problem Description
|
|
Nt821731c -stopcock between burette and patient broke.No injuries.
|
| |
|
Event or Problem Description
|
|
(b)(6) -stopcock between burette and patient broke.No injuries.
|
| |
|
Additional Manufacturer Narrative
|
|
Follow up report ref to natus complaint#(b)(4).The lot number of the affected device was not provided by the customer.Risk management review: a review of (b)(4) medical device hazard analysis (rev 13), identifies the hazard situation as "4.9 - the stopcock at the bottom of the burette tube fails during the operation (unable to turn the stopcock valve and/or breakage of the valve)." the risk level is considered moderate.Based on complaint of "broken stopcock handle." this determination is based on the information received from the customer.Related to capa005781.Root cause/failure investigation: the customer did not return the device for evaluation; an image was provided and attached to the complaint record.Additional testing and evaluation could not be performed as the product was not returned.Devices are fully tested prior to release of product.No further action is required; complaints will be tracked and trended for recurring issues.Failure mode: broken / damaged - stopcock.*confirmed by image only.
|
| |
|
Search Alerts/Recalls
|